Round robin tests XRF-analysis nanoanalytics - Quantitative X-ray microanalysis by SEM/EDX - 2019
Study organized by:
                    NanoAnalytics, Münster
                    
                    Material of proficiency test:
                    Laboratory comparison EDX (LV19) - EQAS test by nanoanalytics, Münster for quantitative X-ray microanalysis using SEM/EDX acc. to ISO 22309 on a chromium-nickel steel
                    
                    Study finished:
                    2019 - 6
                    
                    Literature:
                    Report on Laboratory Comparison EDX (LV19)
                major elements [wt %]
| CRB | RV | 1sRV | Z-Score | |
| Fe | 65,200 | 65,700 | 0,770 | -0,600 | 
| Cr | 18,300 | 17,600 | 0,620 | 1,150 | 
| Ni | 11,200 | 12,200 | 0,410 | -2,340 | 
| Mo | 2,850 | 2,610 | 0,330 | 0,730 | 
| Mn | 1,370 | 1,260 | 0,370 | 0,300 | 
| Si | 0,430 | 0,410 | 0,100 | 0,200 | 
| Cu | 0,180 | 0,270 | 0,090 | -1,000 | 
| V | 0,090 | 0,130 | 0,100 | -0,400 | 
| Co | 0,350 | 0,600 | 0,300 | -0,830 | 
trace elements [µg/g]
| CRB | RV | 1sRV | Z-Score | |
Legend
CRB
                            results CRB
                            cooperative test 
                            results collaborative study
                            1s-RV
                            Standard deviation collaborative study
                            Z-Score
                            Difference of measured value from mean study
                            *
                            Value not certified