Home > QM, accreditation, proficiency tests > Round robin tests XRF-analysis > 160-2018-geopt43-ads-1
Logo

Round robin tests XRF-analysis XRF Round-robin test GeoPT 43, England - ADS-1, Dolerite

Study organized by:
International Association of Geoanalysts and Geostandards Newsletter - GeoPT43
Material of proficiency test:
ADS-1, Dolerite
Study finished:
2018 - 6
Literature:
Report - GeoPT43 Proficiency Testing Round 43 (Laborcode CRB = B100)

major elements [wt %]

CRBRV1sRVZ-Score
Na₂O2,4502,4700,031-0,230
MgO5,5605,5900,150-0,410
Al₂O₃13,73013,7400,187-0,030
SiO₂51,02051,0100,4360,010
P₂O50,2960,3000,012-0,280
K₂O0,9900,9990,028-0,240
CaO9,0308,9700,1660,230
TiO₂2,4302,4300,0490,000
Fe₂O₃tot13,29013,2200,2570,200
MnO0,1730,1710,0050,260

trace elements [µg/g]

CRBRV1sRVZ-Score
Ba359,00341,6022,700,76
Ce40,0059,503,90-3,79
Co52,0044,604,101,82
Cr118,0091,8012,203,52
Cu83,0064,305,703,40
Ga22,0021,901,700,05
Hf4,505,100,30-0,94
La30,0026,902,201,17
Nb15,0017,401,50-1,34
Nd29,0032,802,00-1,21
Ni54,0052,804,900,27
Pr8,007,800,400,27
Rb22,0025,301,50-1,32
Sc28,0027,602,400,17
Sr409,00402,0015,700,27
V327,00333,0019,90-0,27
Y30,0030,002,200,00
Zn135,00113,009,802,48
Zr210,00199,0010,500,77

Legend

CRB
results CRB
cooperative test
results collaborative study
1s-RV
standard deviation collaborative study
Z-Score
Difference of measured value from mean study
*
Value not certified

Asbestos Analysis0

Reliable analysis of asbestos with SEM/EDXORDER NOW!

Order inquiry & contact

  • Phone: +49 (0)5505 // 940 98-0
  • fax: +49 (0)5505 // 940 98-260

News

Apr 16, 2019

Meet CRB at the 8th symposium "Der Bausachverständige"

Our experts welcome you at the 8th symposium "Der Bausachverständige" (The Building Expert) in Cologne on June 5. The focus of this year's event is on indoor pollutants radon and asbestos. ... read more

Mar 4, 2019

Test method developed for asbestos decontamination of firefighters' clothing

In fire situations, firefighters are exposed to a variety of hazardous substances, including asbestos. In cooperation with the MeyerundKuhl specialist laundry, CRB has developed a test procedure to check whether fire-fighting textiles can be sufficiently decontaminated by washing and impregnation. ... read more

Feb 20, 2019

Review of the DCONex 2019: Building pollutants & construction in existing buildings

We report on our impressions of DCONex 2019, which took place on 30 and 31 January in Essen. For the first time we were represented at the fair with our own stand. ... read more