Home > QM, accreditation, proficiency tests > Round robin tests XRF-analysis > 104-2015-flux-zement-121
Logo

Round robin tests XRF-analysis XRF Round-robin test Fluxana, Germany - Cement FLX-CRM 121

Study organized by:
FluXana GmbH & Co. KG, Borschelstr. 3, 47551 Bedburg-Hau
Material of proficiency test:
FLX-CRM 121
Study finished:
2015 - 3
Literature:
Zertifikat Fluxana

major elements [wt %]

CRBRV1sRVZ-Score
Al₂O₃8,8308,7820,2820,200
CaO44,81044,7550,6640,100
Fe₂O₃3,0103,0750,107-0,700
K₂O2,7202,7120,1290,000
MgO1,4101,4050,0820,100
Na₂O1,2701,2220,1110,400
P₂O50,1120,1130,008-0,100
SiO₂30,87030,7430,5480,200
SO₃2,7902,5260,1172,200
TiO₂0,3530,3510,0190,100
LOI3,6704,1480,147-3,300

trace elements [µg/g]

CRBRV1sRVZ-Score
Cr₂O₃0,0080,0090,004-0,20
Mn₂O₃0,1850,1820,0160,00
SrO0,1400,1370,0140,30
ZnO0,0290,0300,006-0,30

Legend

CRB
results CRB
cooperative test
results collaborative study
1s-RV
standard deviation collaborative study
Z-Score
Difference of measured value from mean study
*
Value not certified

Asbestos Analysis0

Reliable analysis of asbestos with SEM/EDXORDER NOW!

Order inquiry & contact

  • Phone: +49 (0)5505 // 940 98-0
  • fax: +49 (0)5505 // 940 98-260

News

Mar 23, 2020

INTERSCHUTZ moved to 2021

The leading trade fair for fire, rescue, civil protection and security will be postponed from this year's June to June 2021 due to the current pandemic situation. ... read more

Mar 19, 2020

Still in service despite Corona

Despite the critical situation, we are still fully operational thanks to targeted internal measures. ... read more

Mar 12, 2020

IFAT 2020 in Munich postponed to September

Due to the continuing spread of the novel coronavirus, IFAT 2020, which was originally scheduled for May 4 to 8, will unfortunately be postponed. ... read more

';