Home > QM, accreditation, proficiency tests > Round robin tests XRF-analysis > 071-2012-flux-zement-106
Logo

Round robin tests XRF-analysis XRF Round-robin test Fluxana, Germany - Cement FLX-CRM 106

Study organized by:
FluXana GmbH & Co. KG, Borschelstr. 3, 47551 Bedburg-Hau
Material of proficiency test:
FLX-CRM 106 - Zement CEM I 42.5 R, Lengerich, Germany
Study finished:
2012 - 2
Literature:
Zertifikat Fluxana

major elements [wt %]

CRBRV1sRVZ-Score
Al₂O₃5,6805,7000,1120,240
CaO65,81066,0500,4400,560
Fe₂O₃1,9501,9760,0780,390
K₂O0,8650,8600,1220,040
MgO0,9900,9620,0510,460
Na₂O0,0700,1240,0441,340
P₂O50,1100,1110,0070,200
SiO₂20,53020,2900,2401,000
SO₃2,9203,0100,1400,650
TiO₂0,2850,2710,0111,220
LOI*1,7402,0600,0803,910

trace elements [µg/g]

CRBRV1sRVZ-Score
Cr₂O₃0,0080,0080,0030,09
Mn₂O₃0,1530,1610,0180,50
ZnO0,0120,0120,0040,07
SrO0,2170,2060,0120,89
Cl0,0310,0550,0350,70

Legend

CRB
results CRB
cooperative test
results collaborative study
1s-RV
Standard deviation collaborative study
Z-Score
Difference of measured value from mean study
*
Value not certified

Asbestos Analysis0

Reliable analysis of asbestos with SEM/EDXORDER NOW!

Order inquiry & contact

  • Tel.: +49 (0)5505 // 940 98-0
  • fax: +49 (0)5505 // 940 98-260

News

Jul 30, 2020

Nachholtermin: vdu-Schulung zum Thema „LAGA PN 98“ im September 2020 in Hannover

Nachholtermin der vdu-Schulung „LAGA PN 98“ anberaumt zum 16. September 2020. vdu-Geschäftstelle in Hannover eruiert Interessenlage und bittet um Mitteilung der Teilnehmerzahl.  ... read more

Mar 23, 2020

INTERSCHUTZ moved to 2021

The leading trade fair for fire, rescue, civil protection and security will be postponed from this year's June to June 2021 due to the current pandemic situation. ... read more

Mar 19, 2020

Still in service despite Corona

Despite the critical situation, we are still fully operational thanks to targeted internal measures. ... read more

';