Accueil > gestion de qualité – accréditation – essais comparatifs > Essais comparatifs SFX > 2001-geopt8-ou-4
Logo

Essais comparatifs SFX XRF Round-robin test GeoPT8 - OU-4, Microdiorite

Organisateur de l'essai interlaboratoires:
International Association of Geoanalysts and Geostandards Newsletter - GeoPT8
Matériel d'essai de l'anneau:
OU-4, Microdiorite
Étude terminée:
2001 - 3
littérature:
Preliminary Report - GeoPT8 Proficiency Testing Round

Éléments principaux [%]

CRBRV1sRVZ-Score
Na₂O3,5603,6100,060
MgO2,2702,3000,041
Al₂O₃14,93014,8300,200
SiO₂63,40063,3400,680
P₂O50,1770,1730,005
SO₃0,020  
K₂O2,7002,7000,047
CaO4,4904,4800,072
TiO₂0,7910,0770,016
Fe₂O₃ tot.5,8205,8600,090
MnO0,1400,1400,004

Les oligo-éléments [µg/g]

CRBRV1sRVZ-Score
Ba355,00360,1012,00
Ce53,0055,702,40
Co16,0013,500,70
Cr56,0054,702,40
Cu26,0027,301,30
Ga18,0017,400,90
Hf4,805,540,34
La32,0025,001,20
Nb11,0012,800,07
Nd27,0027,901,35
Ni25,0021,000,76
Pb18,0014,100,76
Pr9,006,850,41
Rb102,0098,503,90
Sr106,0099,904,00
Th9,008,420,49
V79,0082,703,40
Y47,0047,102,10
Zn71,0069,502,90
Zr197,00195,107,10

légende

CRB
Les résultats CRB
essai interlaboratoires
Résultats collaborative study
1s-RV
écart-type étude collaborative
Z-Score
Différence entre la valeur mesurée et l'étude moyenne
*
Valeur non certifié

Analyse d’amiante0

Analyse d’amiante fiable moyennant la microscopie électronique à balayage, MEB / EDXCOMMANDEZ MAINTENANT !

Demande et le contact

  • Tél.: +49 (0)5505 - 940 98-0
  • fax: +49 (0)5505 - 940 98-260

Envoi d'échantillon

Instructions pour le prélèvement et l'expédition de l'échantillon