Home > QM, Acreditación, EQAS > Análisis de fluorescencia de rayos X EQAS > 055-2009-geopt-24-ou-10
Logo

Round robin prueba el análisis XRF XRF EQAS Geopt24 - OU-10, color gris

VAT No.:
Boletín de la Asociación Internacional de Geoanalistas y Geoestándares - Geopt24
Material de comparación entre laboratorios:
OU-10, Longmydian Greywacke
Estudio terminado:
2009 - 2
Bibliografía:
Informe de Pruebas de Competencia GeoPT24 (Laborcide CRB = Y20)

elementos principales [wt %]

CRBRV1sRVZ-Score
Na₂O2,4302,3100,043
MgO1,7401,7700,033
Al₂O₃11,02010,9700,153
SiO₂73,44073,1200,764
P₂O50,0890,0900,003 
K₂O1,2601,2800,025
CaO2,3302,3400,041
TiO₂0,5200,5340,012
Fe₂O₃ tot4,9004,9200,077
MnO0,1210,1200,003
L.O.I.2,1102,2000,039

oligoelementos [µg/g]

CRBRV1sRVZ-Score
As2,702,80 
Ba331,00311,0010,50
Ce22,0038,001,80
Co13,0012,000,70
Cr36,0034,001,60
Cs1,601,700,12
Cu21,0022,301,10
Ga10,0012,000,70
Ge0,801,40 
Hf3,303,300,20
Hg0,300,35 
In0,030,04 
La11,0018,801,00
Mo4,000,980,08
Nb7,007,700,40
Nd15,0018,701,00
Ni15,0017,700,90
Pb24,0026,901,30
Pr3,004,700,30
Rb33,0036,001,70
Sb3,701,20 
Sm2,903,900,30
Sr166,00174,006,40
Ta135,00124,004,80
Te0,050,06 
Th4,705,000,30
Tl0,250,23 
V77,0077,003,20
W2,902,20 
Y18,0020,501,00
Zn52,0054,002,40
Zr124,00123,004,80

Leyenda

CRB
resultados CRB
essai interlaboratorios
Resultados del estudio colaborativo
1s-RV
Estudio colaborativo de desviación estándar
Z-Score
Diferencia entre el valor medido y el estudio promedio
*
Valor no certificado

Análisis del amianto0

Análisis fiable del amianto mediante microscopía electrónica de barrido, MEB/EDX¡PIDA AHORA!

¡Le asesoro con mucho gusto!

  • Tel.: +49 (0) 5505 // 940 98-0
  • Fax: +49 (0) 5505 // 940 98-260

Noticias

29/11/2022

Follow-up report: 31st Forum Asbestos and other Building Pollutants

For the 31st time, the Forum Asbestos took place at the Haus der Technik in Essen on November 10th and 11th, 2022. In order to do justice to the continuing situation, this year there was the possibility to participate in person on site, as well as to join online. Here is a summary of our impressions of the conference: Leer más sobre

13/10/2022

New vdu trainings until the end of 2022

By the end of 2022, the Association of German Testing Laboratories (vdu) would like to hold training at the vdu office in Hannover. As a member of the vdu, we are pleased to point out this interesting opportunity.  Leer más sobre

16/9/2022

La limpieza incorrecta de tejados con amianto conlleva penas de cárcel y multas

Es necesario tener especial precaución cuando se trata de amianto. El polvo de amianto puede liberarse no sólo durante los trabajos de renovación, sino también cuando los tejados de amianto se limpian de forma inadecuada. Los que hacen caso omiso de la normativa se ponen en peligro a sí mismos y a sus compañeros y se arriesgan a recibir fuertes multas. Leer más sobre